that use two or three gates depending on the multiple echo mode selected, echo-
echo (E-E) or echo-echo verify (E-EV).
The procedures to adjust the Thresholds are outlined below:
1) Press the MEAS key once to activate measure menu items. Press the
MEAS key multiple times to move down and the ESC key multiple times to
move up, until the desired Threshold cell is highlighted.
2) Press the
button wheel to scroll the highlighted value.
3) Alternatively, press the ENTER key to display the Digits Edit Box.
4) Press the
right button wheel to scroll the highlighted value.
5) Press the
right button wheel to scroll the digit locations.
6) Repeat steps 4 & 5 until the Threshold value is correctly displayed.
7) Press the OK key to return to the measurement screen, or ESC to cancel
entering the Threshold.
The user can also access and adjust the Threshold from the tabbed menus.
However, this method is more tedious than making the adjustments using the Hot
Menus. The procedure using the tabbed menus is outlined below: