13.9 Damping

The MAX II has a built-in damping feature to control the impedance input of the
receiver. This enables the user the ability to match and optimize the transducer for
better signal quality at various frequencies. The available settings are 50, 75, 100,
300, 600 and 1500 ohms. The procedures below outline the steps for selecting an
impedance setting, as follows:
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right, and the ESC key multiple times to tab left,
until the TUNE menu is highlighted and displaying the submenu items.
Alternatively, press the
using the left button wheel.
2) Use the
wheel to scroll through the sub menu items until DAMPING is highlighted.
Find Quality Products Online at:
LEFT and RIGHT arrow keys multiple times on the right button
Setting the Damping Value
LEFT and RIGHT arrow keys multiple times
UP and DOWN arrow keys multiple times using the left button