DAKOTA ULTRASONICS MAX II Operation Manual page 69

Ultrasonic bolt tension monitor
Hide thumbs Also See for MAX II:
Table of Contents

Advertisement

Dakota Ultrasonics
menu items. Alternatively, press the
multiple times using the left button wheel.
7) Press the
button wheel to scroll through the sub menu items until APPROX LEN is
highlighted.
8) Press the ENTER key to display the Digits Edit Box and adjust the
APPROX LEN for 2" to find the correct length.
9) Press the
the left or right button wheel to scroll the highlighted value until the desired
VELOCITY is correctly displayed.
10) Press the OK key to save the APPROX LEN and display the AUTO menu
items.
Note: Proceed to the verification section as a guide to access and open the
CAL_ZERO_DATA project, CALZERO data group, and record the 2" side
of the glass block. Follow steps 1 – 9, The procedure for recording the 2"
point will be at the 'end' of the verification section. However, start from the
beginning to open the group data file.
Caution:
A very small amount of "WATER", or other non-intrusive fluid with a similar viscosity,
should be used as the couplant between the transducer and calibration standard
selected. This will avoid creating a couplant film layer between them and cause an
Find Quality Products Online at:
UP and DOWN arrow keys multiple times using the left
UP, DOWN, LEFT, & RIGHT arrow keys multiple times using
64
GlobalTestSupply
www.
LEFT and RIGHT arrow keys
.com
sales@GlobalTestSupply.com

Advertisement

Table of Contents
loading

Table of Contents