As stated above, the MAX II has a total of three gates depending on the
measurement mode selected. The photo above illustrates the mode and number of
gates needed for each measurement mode. One gate is used in pulse-echo, two
gates for echo-echo, and 3 gates for echo-echo verify. The gates are fully
independent, and starting positions and thresholds are the only adjustments available
for each gate.
1) Press the MEAS key once to activate measure menu items. Press the
MEAS key multiple times to move down and the ESC key multiple times to
move up until the GATE cell is highlighted. Alternatively, press the
and DOWN arrow keys multiple times using the left button wheel.
2) Press the
button wheel to scroll the highlighted value.
3) Alternatively, press the ENTER key to display the Digits Edit Box.
Find Quality Products Online at:
Measurement Modes & Gates
Adjusting the Gate using the Hot Menus
LEFT and RIGHT arrow keys multiple times using the left