18) Press the
button wheel to scroll through the sub menu items until POLARITY is
highlighted.
19) Press the
button wheel to scroll the POLARITY options (auto, positive, negative) until
the desired option is displayed.
Note: AUTO is the default option and will analyze both sides of the waveform
(positive/negative) and attempt to determine the optimal polarity.
Experienced operators can opt to override this option by directly setting the
polarity to either POSITIVE or NEGATIVE as is needed.
20) Clean both the cal bar and transducer surface and apply a small drop of
couplant to the surface of the cal bar. Attached the transducer and rotate it
clockwise and counter clockwise to force out the excess couplant. Center
the transducer. Be methodical with this process to eliminate any potential
errors.