DAKOTA ULTRASONICS MAX II Operation Manual page 61

Ultrasonic bolt tension monitor
Hide thumbs Also See for MAX II:
Table of Contents

Advertisement

Dakota Ultrasonics
5) Press the ENTER key to display the Zero Calibration menu items.
6) Press the
button wheel to scroll through the sub menu items until PHY LEN 1 is
highlighted.
7) Press the ENTER key to display the Digits Edit Box and adjust the physical
length of the cal bar.
8) Press the
the left or right button wheel to scroll the highlighted value until the desired
PHY LEN 1 is correctly displayed.
Note: PHY LEN 1 is the actual physical length of the calibration standard noted
on the calibration certificate.
9) Press the OK key to save PHY LEN 1 and display the Zero Calibration
menu items.