DAKOTA ULTRASONICS MAX II Operation Manual page 73

Ultrasonic bolt tension monitor
Hide thumbs Also See for MAX II:
Table of Contents

Advertisement

Dakota Ultrasonics
1) Press the
the left button wheel to scroll the cursor to an empty L-REF cell in column A,
row #4 – 6.
Note: Clean the surface of the 2" side of the Glass Block thoroughly. Apply a
small drop of WATER, or other non-intrusive fluid with a similar viscosity, in
the center of the 2" side and attach the transducer to the Glass Block.
2) Press the AUTO SET key to find the 2" echo and optimize the MAX II setup.
Note: Review the waveform and length displayed for a tolerance 0.0005".
Rotate, maneuver, and vary pressure on the transducer, while searching for
a length that's inside of allowable tolerance.
3) Press the ENTER key to store/record the 2" side of the glass block.
4) Press the MENU key to display the DATA sub menu items.
5) Press the
button wheel to scroll through the sub menu items until CLOSE is
highlighted.
6) Press the ENTER key to close the CALZERO group, followed by pressing
the MEAS key to return the main measurement screen.