Dakota Ultrasonics
Note: The system zero calibration routine just performed is entirely
independent of the bolt measurement section. Therefore, be sure the MAX
II is setup for the actual bolts intended prior to measurement.

8.6 Verification of System Zero Calibration

The system zero can be verified either immediately following completion of one of the
zero calibration methods described in the previous sections, prior to starting a
new/existing job, or because the hardware has been changed in some way (gauge,
transducer, or cable).
Note: If the glass block was used in conjunction with the two point calibration
method, with the intension of using all three points (1", 2", 3"), recordation of the 2"
side of the block can also be stored/documented using this procedure to access the
CALZERO group data file. However, in order achieve an accurate result, the velocity
calculated at the end of the zero calibration procedure must be entered as the
velocity in the MATL menu. The zero calibration section is independent of the 'live'
measurement section. Additionally, the approximate length must changed to capture
the 2" side of the glass block.
The procedure below will guide the operator through changing the velocity &
approximate length prior to opening the CALZERO group data file:
1) Press the MENU key multiple times to tab right and the ESC key multiple
times to tab left until the MATL menu is highlighted and displaying the sub