21) Press the
button wheel to scroll through the sub menu items until ULTRASONIC 1 is
highlighted.
22) Press the ENTER key to start the automated zero calibration routine.
Note: A waveform will be displayed with "CALC ZERO", noted in the top right
corner of the display. Once the routine is completed and a stable echo is
displayed, review the waveform quality and polarity selected.
23) Press the ENTER key to store the result - ACCEPT CAL.
24) Repeat steps 20 – 23 if using the TWO POINT zero calibration option, to
complete the routine for the additional zero cal bar ULTRASONIC 2.
Note: The operator can run the ultrasonic routine as many times as is needed,
or until comfortable with the results. When satisfied with the results, it's
time to calculate the zero.
25) Press the
button wheel to scroll through the sub menu items until CALC ZERO is
highlighted.