perhaps a 50' press bar, the high voltage (HV) and tone burst (TB) options are
available. The procedure to change the pulse width is outlined below:
For quick access and editing, refer to navigating the Hot Menus page 36.
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple times to tab left until
the TUNE menu is highlighted and displaying the submenu items.
Alternatively, press the
using the left button wheel.
2) Use the
wheel to scroll through the sub menu items until PULSE is highlighted.
3) Use the
wheel to scroll the Pulse options.
4) Once the Pulse is displayed, press the MEAS key to return to the
measurement screen.

13.8 Pulser Voltage

The MAX II has a 400 volt square wave pulser that can be adjusted for specific bolt
lengths, geometries, materials and transducer size and frequencies. The Pulser
Voltage feature offers various voltage settings of 100 to 400 volts, based on the pulse
width setting. Lower voltage selections are best for short small diameter bolts, while
higher voltage are best for longer bolts with attenuative or inconsistent materials.
Find Quality Products Online at:
Selecting the Pulse
LEFT and RIGHT arrow keys multiple times
UP and DOWN arrow keys multiple times using the left button
LEFT and RIGHT arrow keys multiple times on the right button
GlobalTestSupply
www.
MAX II Bolt Tension Monitor
143
.com
sales@GlobalTestSupply.com

Advertisement

Table of Contents
loading

Table of Contents