DAKOTA ULTRASONICS MAX II Operation Manual page 105

Ultrasonic bolt tension monitor
Hide thumbs Also See for MAX II:
Table of Contents

Advertisement

Dakota Ultrasonics
1) Press the MENU key multiple times to tab right and the ESC key multiple
times to tab left until the CAL menu is highlighted and displaying the
submenu items. Alternatively, press the
multiple times using the left button wheel.
2) Press the
button wheel to scroll through the sub menu items until LOAD CAL MODE
is highlighted.
3) Press the
button wheel to scroll through the unit options OFF, VECTOR (zero offset),
REGRESSION, until the correct mode is displayed.
4) Press the MEAS key to return to the measurement mode.
Now that a group has been created and activated, and the load calibration mode
selected, we're ready to perform the field calibration.
1) Place the first sample bolt in the load apparatus.
Note: The following steps assume that the user has previously selected a
similar material type for the list of materials and the approximate length of the
bolt has been entered. Therefore, the MAX II is currently setup and ready to
measure in elongation mode.
Find Quality Products Online at:
UP and DOWN arrow keys multiple times using the left
LEFT and RIGHT arrow keys multiple times using the right
Performing a Field Calibration
100
GlobalTestSupply
www.
LEFT and RIGHT arrow keys
.com