13.7 Pulse

The MAX II has a very powerful 400 volt square wave and 400 volt tone burst pulser
that can be adjusted using the pulse width option. Pulse width, refers to the duration
of time the pulser is left on. This time results in increased energy sent into the test
material. There are three square wave pulse width options (SPIKE, THIN, and
WIDE) with adjustable pulser voltages of 100, 150 and 200 volts. There are an
additional three high voltage square wave options (HV SPK, HV THIN, and HV
WIDE) with adjustable pulser voltages of 200, 300 and 400 volts.
The there are additionally five tone burst pulser options (TB10MHZ, TB 5MHZ, TB
2MHZ, TB 1 MHZ, and TB.5MHZ) with adjustable pulser voltages of 200, 300 and
400 volts. The SPIKE setting may be desirable for high resolution and general
applications to decrease the overall noise. When additional energy is required for
142