DAKOTA ULTRASONICS MAX II Operation Manual page 56

Ultrasonic bolt tension monitor
Hide thumbs Also See for MAX II:
Table of Contents

Advertisement

Caution:
A very small amount of "WATER", or other non-intrusive fluid with a similar viscosity,
should be used as the couplant between the transducer and calibration standard
selected. This will avoid creating a couplant film layer between them and cause an
"out of tolerance" condition > +/- 0.001" for the 3" cal bar.
1) Press the MENU key multiple times to tab right and the ESC key multiple
times to tab left until the CAL menu is highlighted and displaying the
submenu items. Alternatively, press the
multiple times using the left button wheel.
2) Press the
button wheel to scroll through the sub menu items until ZERO MODE is
highlighted.
3) Press the
button wheel to scroll the zero mode options until the AUTO option is
displayed.
Note: The auto option will perform a two-point calibration using a single 3" cal
bar (pt# X-000-0011) and measurements of 1x & 2x lengths.