Power Off Settings - Konica Minolta mKDR Xpress Operation

Digital mobile x-ray unit
Table of Contents

Advertisement

3.2.9

POWER OFF SETTINGS

.
Note
OM-0481R1
Power OFF settings can be configured by the User Administrator
or the Service Engineer.
" Time Out Switch Off Pressed. Time set by default:: 3 seconds.
Press and hold the Switch OFF button for 3 seconds; the system will start the
turning OFF process.
" Time Out Delay Switch Off. Time set by default:: 50 seconds.
After holding pressed for 3 seconds the Switch OFF button, the system will be
completely turned OFF in 50 seconds; during this time only movements are
enabled (e. g. it is useful for the user to put the system in the Parking position,
if necessary).
" Time Out Inactivity Generator. Time set by default:: 2 minutes.
When the system is disconnected from mains and there is not any Generator
activity during 2 minutes, the Generator Tube Filaments turn off, unless the user
performs any action on the RAD Screen, presses any control of the Handswitch
or Infrared Remote Handswitch, or connects the system to the mains; it will
delay the Time Out Inactivity Generator for another 2 minutes without any
Generator activity.
" Time Out Inactivity Warning. Time Set by default: 15 minutes.
When the system is disconnected from the mains and there is not any
Generator activity for 15 minutes, a Warning message is displayed on the
screens to alert the user of an automatic shutdown after finishing a countdown
timer for another 15 minutes.
If the Operator presses on Accept button or performs any system action
(touches on the RAD Screen, activates any of the movement controls, presses
the Handswitch or Infrared Remote Handswitch, or connects the system to the
mains), within the configured time; it will delay the Time Out Inactivity Warning
for another 15 minutes.
If the Operator does not respond within the countdown time of 15 minutes, the
system will automatically shut down; that is, the system will be completely
turned OFF after 30 minutes without activity.
Mobile X-ray Unit