2.7.4

PROTECTION AGAINST ELECTRIC SHOCK HAZARDS

2.7.5

PROTECTION AGAINST HARMFUL INGRESS OF WATER OR PARTICULATE MATTER

2.7.6

PROTECTION AGAINST HAZARDS OF IGNITION OF FLAMMABLE ANAESTHETIC

MIXTURES

2.7.7

PROTECTION AGAINST HAZARDS FROM UNWANTED OR EXCESSIVE RADIATION

OM-0481R1
Protection against electric shock hazards in accordance with Standards:
IEC 60601-1:1988, IEC 60601-1:2005 and IEC 60601-1:2005+AMD1:2012,
IEC 60601-2-54:2009 and IEC 60601-2-54:2009+AMD1:2015.
This equipment has been classified as a type-B ( ) device, in accordance with
Standard IEC 60601-1 requirements: Class I - Type B applied parts.
TO AVOID THE RISK OF ELECTRIC SHOCK, THIS
EQUIPMENT MUST ONLY BE CONNECTED TO A SUPPLY
MAINS WITH PROTECTIVE EARTH.
Protection against harmful ingress of water or particulate matter: Ordinary
(IPx0), in accordance with Standard IEC 60601-1:1988, IEC 60601-1: 2005
and IEC 60601-1:2005+AMD1:2012.
Degree of Safety in the presence of Flammable Anesthetics Mixture with air or
with oxygen or with nitrous oxide: Not suitable for use in the presence of
Flammable Anesthetics Mixture with air or with oxygen or with nitrous oxide, in
accordance with Standard IEC 60601-1:1988,
IEC 60601-1:2005+AMD1:2012.
Protection against hazards from unwanted or excessive radiation in
accordance with Standards
IEC 60601-1:2005+AMD1:2012
60601-1-3:2008 and IEC 60601-1-3:2008+AMD1:2013.
IEC 60601-1: 2005 and
IEC 60601-1:1988, IEC 60601-1: 2005 and
and
IEC60601-1-3:1994,