3.2.8

BATTERY STATE ALERTS

.
Note
OM-0481R1
The following functions may be configured by the Service
Engineer.
For further information, refer to Section 4 of Control Console and
Section 5 of System Messages.
The Following Battery State Alerts can be displayed on the RAD screen:
" Battery Low:
This alert appears when the system batteries are at Low Charge status.
The lower indicators light solid green and the middle and upper indicators are
off, with the unit disconnected from mains (not charging).
It indicates that normal operation is allowed although it is recommended to
connect the unit to the power supply, for better battery usage.
The message disappears after a few seconds or by clicking on the OK button;
then, the work can be continued.
" Battery Very Low:
This alert appears when the system batteries are at Very Low Charge status.
The lower indicators light solid orange and the middle and upper indicators are
off, with the unit disconnected from mains (not charging). It indicates that normal
operation is allowed although it is urgent to connect the unit to the power supply.
The message disappears after a few seconds or by clicking on the OK button;
then, the work can be continued.