Konica Minolta mKDR Xpress Operation page 168

Digital mobile x-ray unit
Table of Contents

Advertisement

Mobile X-ray Unit
Operation
C-14
" POWER OFF SETTINGS
Time Out Switch Off Pressed.
Default configured time: 3 seconds.
Press and hold the Switch OFF button for 3 seconds; the system will start the
turning OFF process.
Time Out Delay Switch Off.
Default configured time: 50 seconds.
After holding pressed for 3 seconds the Switch OFF button, the system will be
completely turned OFF in 50 seconds; during this time only movements are
enabled (e.g., it is useful for the user to put the system in the Parking position,
if necessary).
Time Out Inactivity Generator.
Default value: 0 (Disabled).
Filaments deactivation due to inactivity is disabled. The following is an example
of how the equipment would behave if the value "2" (minutes) were entered:
When the system is disconnected from mains and there is not any
Generator activity during 2 minutes, the Generator Tube Filaments turn
off, unless the user performs any action on the RAD screen, presses the
handswitch/remote control (option) or connects the system to the mains;
it will delay the "Time Out Inactivity Generator" for another 2 minutes
without any Generator activity.
Time Out Inactivity Warning.
Default value: 0 (Disabled).
No system shutdown due to inactivity. The following is an example of how the
equipment would behave if the value "15" (minutes) were entered:
When the system is disconnected from the mains and there is not any
Generator activity for 15 minutes, a Warning message is displayed on the
screens to alert the user of an automatic shutdown after finishing a
countdown timer for another 15 minutes.
Appendix C