PERIODIC MAINTENANCE

134
Do not allow the batteries to be deeply discharged because they will lose
storage capacity and will never be able to recover the 100% of their
original capacity.
For more information, refer to "Battery Charge Level Indicators" in
Section 3.2.7 and "Battery Capacity for the Generator and the
Motors" in Section 8.1.
The first periodic maintenance service should be performed six (6) months after
installation, and the subsequent services at twelve (12) month intervals.
Periodic maintenance tasks shall include the following items:
DO NOT REMOVE ANY COVER, DISASSEMBLE OR
MANIPULATE INTERNAL COMPONENTS IN THE UNIT.
THESE ACTIONS COULD CAUSE SERIOUS PERSONAL
INJURIES AND / OR EQUIPMENT DAMAGE.
1.
With the Unit OFF, plug it in and leave it sufficient time to completely
charge. The recommended time is approximately 8 hours, until the
Battery Charge Level Indicators on both columns stop scrolling and all
the Indicators light Green.
2.
Once fully charged, unplug the Unit from the mains power. Wait a few
minutes and reconnect the Unit to the mains. The upper Green Indicators
should scroll up for approximately one minute.
If the Battery charge level Indicators begin to scroll up from any other
Indicator below, contact the Service Department.
3.
Switch the equipment OFF by shutting down the computer. Turn OFF the
switch ON/OFF control and unplug the unit from mains.
4.
Check the external cable connections.
OM-0481R1