3.6.1.2

FINE POSITIONING CONTROLS

Fine Positioning Controls:
1 Backward
2 Forward
.
Note
OM-0481R1
Fine Positioning Controls permit the fine positioning adjustment of the system
respecting the patient, with the operator positioned opposite the
Head-Assembly.
The buttons with the arrow icons located on each Handgrip, control the motion
of each driving wheel independently, at low speed, backward/forward.
The buttons correspond to each motor and operate with the unit in/out of the
Parking Position.
Once any Fine Positioning button is pressed, the Unit starts to move, but
displacement will stop after 5 seconds. It is necessary to release the button and
press again to resume the movement. This safety measure prevents possible
uncontrolled movements due to an internal failure of the Fine Positioning
Controls.
Fine Positioning velocity is reduced as this control is not designed for
displacements.
1
2
For further information on Fine Positioning Controls, refer
to Illustration 3-3.
In order to avoid uncontrolled displacement of the unit during
the Start-up, due to a failure of the displacement controls
(Fine Positioning Controls activated or short-circuited),
movements controlled with these commands are blocked
although the unit can be controlled with the Handlebar.
A warning message alerts the user about a failure condition.
Mobile X-ray Unit
Operation