Table 3-1
Image Size according to the SID and Collimator Opening
OM-0481R1
The Collimator has the following functions:
1.
Collimator LED Light push-button.
Press once the Collimator Light push-button to turn the light ON; it will
remain ON for a few seconds before automatically switching OFF.
Collimator light may also be switched ON/OFF by clicking the Collimator
Light push-button located at the Handswitch.
Press two times within 5 seconds to turn ON the Double Laser in order
to adjust the SID.
Press three times to turn OFF the light (if it remains ON).
2.
Measuring tape.
To measure the SID.
3.
Handgrips with Support .
For easily positioning the Head-Assembly.
4.
Collimator knobs to adjust the internal blades.
The Exposure Field is adjusted by setting the Knobs in order to
open/close the internal blades.
COLLIMATOR
COLLIMATOR
OPENING
13
15
18
24
30
35
40
43
90 cm (35.4")
100 cm (39.4")
11.7 cm (4.6")
13 cm (5.1")
13.5 cm (5.3")
15 cm (5.9")
16.2 cm (6.4")
18 cm (7.1")
21.6 cm (8.5")
24 cm (9.4")
27 cm (10.6")
30 cm (11.8")
31.5 cm (12.4")
35 cm (13.8")
36 cm (14.2")
40 cm (15.7")
38.7 cm (15.2")
43 cm (16.9")
Mobile X-ray Unit
Operation
SID
180 cm (70.9")
23.4 cm (9.2")
27 cm (10.6")
32.4 cm (12.8")
43.2 cm (17")
54 cm (21.3")
63 cm (24.8")
72 cm (28.3")
77.4 cm (30.5")