Konica Minolta mKDR Xpress Operation page 126

Digital mobile x-ray unit
Table of Contents

Advertisement

Mobile X-ray Unit
Operation
ID
140608
Generator Starter. Voltage Warning.
140609
Generator Starter. Voltage Warning.
140610
Generator Starter. Voltage Warning.
Generator Starter. Tube Impedance Read
140611
Information.
140612
Generator Starter Interlock.
Generator Starter Interlock
140613
140616
Generator Starter powered Off
290001
System Communication Error with GPIO Head
290002
System Communication Error with Battery Monitor
290003
System Communication Error with RFID
290004
System Communication Error with ADMC
290005
290006
Component Identification Mismatch
Component Identification Mismatch
290007
290008
290020
Collimator Apertures Error
290100
Positioner configuration checksum is wrong
290101
Configuration file not loaded
290102
Configuration file not loaded
118
DESCRIPTION
TYPE
Design Information - Generator Starter.
Information
V_ACC Maximum Duty Warning.
Design Information - Generator Starter. V_HOLD
Information
Maximum Duty Warning.
Design Information - Generator Starter.
Information
V_DC Maximum Duty Warning.
Design Information - Generator Starter. Tube
Information
Impedance Read Value.
Error
Error
Wait until the message dissapears.
Wait until the message dissapears
Information
No user action required.
Restart the System.
Error
If the error persists, call Service.
Restart the System and try to repeat the operation.
Error
If the error persists, call Service.
If the error persists, call Service.
Error
Error
Call Service.
Call Service
Warning
Call Service.
Restart the System and try to repeat the operation.
Error
If the error persists, call Service.
Restart the System.
y
E
Error
If the error persists, call Service.
USER HELPTEXT
OM-0481R1

Advertisement

Table of Contents
loading