Konica Minolta mKDR Xpress Operation page 117

Digital mobile x-ray unit
Table of Contents

Advertisement

ID
DESCRIPTION
101022
Generator inactivity state, preheating filament
Upgrade is ongoing, xray disabled during this
101023
process
KVM button has been pressed, switching between
101024
computers
The exposure selected exceeds the current battery
101025
capacity power. Exposures are not allowed in this
state.
Upgrading is performing, when the process finishes
101026
click on accept to apply the changes.
102001
Value requested exceeds generator power
102002
Value requested exceeds tube maximum rating
Technique requested not allowed due to tube space
102003
charge
102004
kVp requested out of range
102005
mAs requested out of range
102006
mA requested out of range
102007
ms requested out of range
Focal spot change not allowed due to mA- -mAs
102008
selection
102009
APR warning
102010
Generator thermal limit
102023
Number of exposure reached. Exposure finished.
102024
Exposure finished due to limit mAs reached.
Exposure finished due to limit energy for serial
102025
radiographic exposures.
103001
Filaments disabled by software
103002
Filaments disabled by hardware
OM-0481R1
TYPE
Information
Wait for 10 seconds.
Information
- -
Inhibit
If KVM has not been pressed, call Service
Exposure
Please, decrease mA and increase ms to keep mAs.
Otherwise, connect the system to power source and
Inhibit
wait.
Exposure
If the error persists, call Service.
Information
- -
Information
No user action required.
Information
No user action required.
Information
No user action required.
I f
Information
ti
Call Service.
C ll S
i
Mobile X-ray Unit
Operation
USER HELPTEXT
109