3.2.7

BATTERY CHARGE LEVEL INDICATORS

.
Note
.
Note
PLUGGED INTO MAINS
System in "OFF" or "ON" position
Both Columns are scrolling as
described in the following Table.
OM-0481R1
The column with the Exposure symbol indicates the charge level of the Batteries
used for radiographic operations (X-ray exposures) and the column with the
Mobile X-Ray unit symbol indicates the charge level of the Batteries used for
the Mobile motion (motors).
When plugged into the mains (with the Emergency Switch-OFF deactivated),
the Batteries automatically charge. The color Indicators on both columns
illuminate and scroll from the current Generator battery charge level to 100%,
until the Batteries are fully charged. During the charging process, both columns
scroll up from the same level.
The Batteries require approximately 8 hours for a fully charge.
To charge the Batteries, there is no need to have the Console
turned ON. When the Batteries are fully charged, the Battery
charge level Indicators on both columns stop scrolling and all the
Indicators remain illuminated in green.
When unplugged from mains, the Batteries discharge independently on their
use (X-ray exposures or motors) since the Mobile is provided with one battery
module.
Upon disconnecting the unit from the mains if the unit has been
connected for a short period of time, or after several exposures or
after one heavy duty exposure, the Batteries need at least
30 seconds to stabilize the charge, after which the correct charge
level is shown on the Indicator.
The Battery Charge Level Indicator can be:
MOBILE unit
MOBILE unit
UNPLUGGED FROM MAINS
System in "OFF" position
Both Columns are OFF.
Mobile X-ray Unit
Operation
System in "ON" position and
Console turned ON
Each Column shows the respective
Battery charge level as described in
the following Table.