Mobile X-ray Unit
Operation
Illustration 4-5
Slider for Value Selection
Min value
.
Note
96
INCREASE / DECREASE: Radiographic technique values are increased or
decreased by changing the value moving the "Slider" position.
When the "Slider" is positioned over a value not allowed, its pointer comes back
to the nearest allowed value, according to the limit of the Tube and the Unit.
Slider for value selection
kV: Selects the X-ray Tube voltage.
mA: Selects the X-ray Tube current, changing the mAs value and
keeping constant the select Exposure Time, whenever possible.
mAs: Selects the exposure in mAs, setting the maximum mA available
for the selected Focal Spot and the respective Exposure Time.
s: Selects the Exposure Time in seconds.
(Refer to Section 8 for Factor ranges)
If after pressing any of these buttons, the technique value is
blocked, it could mean that it may have been selected a wrong
combination of radiographic parameters that could have caused
a warning condition (for further information about System
Messages, refer to Section 5).
Max value
OM-0481R1