Konica Minolta mKDR Xpress Operation page 66

Digital mobile x-ray unit
Table of Contents

Advertisement

Mobile X-ray Unit
Operation
MOBILE unit IN CHARGING MODE
(PLUGGED TO MAINS)
LED INDICATORS AND STATUS
Full Charge:
After a complete charge of the system (during approx.
8 hours), the upper indicators on both columns (1+1) light
blinking green for a few seconds and then, all the
Indicators (3+3) light steady green, indicating that the
system Batteries are at Full Charge status.
The batteries charge level is at 100% of the total charge.
Medium Charge:
The lower 4 indicators on both columns (2+2) light steady
green and the upper 2 indicators on both columns (1+1)
light blinking green and scrolling upwards, indicating that
the system Batteries are at Medium Charge status.
Low Charge:
The lower 2 indicators on both columns (1+1) light steady
green and the upper 4 indicators light blinking green and
scrolling upwards, indicating that the system Batteries are
at Low Charge status.
Very Low Charge:
All the indicators on both columns (3+3) are blinking green
and scrolling upwards, indicating that the system
Batteries are at Very Low Charge status.
Critically Low Charge:
All the indicators on both columns (3+3) are blinking green
and scrolling upwards, indicating that the system
Batteries are at Critically Charge status.
Indicator colors:
58
Both columns comprise three Indicators, representing a battery status as
described below: Battery Full, Medium, Low, Very Low and Critically Low.
Green
Orange
MOBILE unit IN STAND-ALONE MODE
(UNPLUGGED FROM MAINS)
LED INDICATORS AND STATUS
When the system is at Full charge status, all the indicators
light steady green and normal operation is allowed.
The batteries charge level is between 60% and 100%
When the system is at Medium Charge status, the lower
indicators light green and normal operation is allowed.
The batteries charge level is between 31% and 59%.
When the system is at Low Charge status, the lower
indicator lights green steady, on one or both columns;
normal operation is allowed although it is recommended
to charge the system.
The batteries charge level is between 20% and 30%.
When the system is at Very Low Charge status, the lower
indicator lights orange steady, on one or both columns;
normal operation is allowed although it is urgent to charge
the system.
The batteries charge level is between 1% and 19%.
When the system is at Critically Low Charge status, the
lower indicator lights blinking orange, on one or both
columns, operation is limited but it is necessary to charge
the system: the indicator on the X-Ray column, indicates
that exposures are disabled; if is on the Motion column,
indicates that motion is enabled for 30 minutes or 1 km
after which, all the indicators will be OFF.
The batteries charge level is 0%.
Indicator OFF