Collimator Movements - Konica Minolta mKDR Xpress Operation

Digital mobile x-ray unit
Table of Contents

Advertisement

3.6.3.3

COLLIMATOR MOVEMENTS

.
Note
OM-0481R1
The Collimator can rotate ¦90_ on its vertical axis (refer to Illustration 3-5,C),
while the Head-Assembly remains in the same position.
This movement is performed by manually turning the Collimator and has
o
o
detents at --90
, 0
and +90
Due to geometric restrictions related to the anode angle of the
X-Ray Tube, a minimum SID is required to cover the full image size
of the Detector, depending on the Collimator position:
X-ray Tube
Anode Angle
Anode Angle
16°
o
.
with Collimator rotated at:
Receptor Size
Receptor Size
0° or
¦
24X30
SID  70 cm
30X24
35X43
43X35
SID  100 cm
SID  100 cm
43X43
Mobile X-ray Unit
Operation
Required SID
90°
45°
¦
SID  70 cm
SID  100 cm
SID  100 cm