SECTION 8
8.1

FACTORS

Maximum Power kW
(Refer to
Identification Label)
kVp Range
kVp Range
mAs Range
mA Range
Exposure Time Range
Focal Spot Size
Power Output (@ 0,1s)
Duty Cycle
Ripple Factor
Collimator
X-ray Tube
OM-0481R1

TECHNICAL SPECIFICATIONS

20 kW
Product of mA x Time values from 0.1 mAs to 500 mAs
10 to 400
From 10 mA to 400 mA or 500 mA, through the following mA stations:
10, 12.5, 16, 20, 25, 32, 40, 50, 63/64, 80, 100, 125, 160, 200, 250, 320, 400, 500.
From 1 millisecond to 10 seconds through the following Time stations:
Milliseconds: 1, 1.25, 1.6, 2, 2.5, 3.2, 4, 5, 6.3/6.4, 8, 10, 12.5, 16, 20, 25, 32, 40,
Seconds:
Maximum Exposure Time Range for DR: From 1 millisecond to 4 seconds.
150 kVp @ 125 mA
125 kVp @ 160 mA
100 kVp @ 200 mA
80 kVp @ 250 mA
62 kVp @ 320 mA
50 kVp @ 400 mA
18 exposures per hour at
100 kV, 200 mA, 320 ms
18 exposures per hour at 120 kV, 250 mA, 250 ms (lapse time between exposure: 3 min.)
(lapse time between
exposure: 3 min.)
Maximum leakage radiation depends on the type of X-ray Tube (<0.88 mGy/h)
Manual Collimator with front and rear side Knobs, double Laser, Motorized Filters
and ready for both Estimated Dose (EDAP) option and DAP Chamber option
32 kW
40 to 150
From 40 kVp to 150 kVp in 1 kVp steps
(Depending on the Generator model)
50, 63/64, 80, 100, 125, 160, 200, 250, 320, 400, 500, 630/640, 800.
1, 1.25, 1.6, 2, 2.5, 3.2, 4, 5, 6.3/6.4, 8, 10.
Small / Large
150 kVp @ 200 mA
128 kVp @ 250 mA
150 kVp @ 250 mA
125 kVp @ 250 mA
125 kVp @ 320 mA
100 kVp @ 320 mA
100 kVp @ 400 mA
80 kVp @ 400 mA
80 kVp @ 500 mA
64 kVp @ 500 mA
< 4% (constant voltage)
Refer to Section 8.2
Mobile X-ray Unit
40 kW
50 kW
10 to 500
150 kVp @ 320 mA
125 kVp @ 400 mA
100 kVp @ 500 mA
Operation
137