Konica Minolta mKDR Xpress Operation page 120

Digital mobile x-ray unit
Table of Contents

Advertisement

Mobile X-ray Unit
Operation
ID
Generator DSP Control. Thermal fault on large
140336
filament inverter.
Generator DSP Control. Undercurrent in super
140340
small filament during exposure.
Generator DSP Control. Undercurrent in the power
converter using super small filament during
140341
exposure.
140342
Generator DSP Control. Overvoltage in super small
Generator DSP Control. Overvoltage in super small
filament.
140343
Generator DSP Control. Following error loop on
140344
super small filament.
Generator DSP Control. IGBT fault on the super
140345
small filament.
Generator DSP Control. Thermal fault on the super
140346
small filament.
140350
Generator DSP Control. Master Heartbeat error.
Generator DSP Control. Emergency signal
140351
activated by master.
Generator DSP Control. Synchronism signal
(exposure order) has been received before
140352
preparation signal.
Generator DSP Control. Filament current demand
140353
is out of range
140354
Generator DSP Control. Frequency out of range.
140355
Generator DSP Control. UDC undervoltage
Generator DSP Control. End exposure not
140356
detected
140357
Generator DSP Control. Internal Status Fail
Generator DSP Control. Monoblock connector
140358
error
140359
Generator DSP Control. mA out of range
112
DESCRIPTION
TYPE
Press Accept.
If the error persists, restart the System.
If the error persists, restart the System.
Error
Error
If the error persists, call Service.
USER HELPTEXT
OM-0481R1

Advertisement

Table of Contents
loading

Table of Contents