Konica Minolta mKDR Xpress Operation page 146

Digital mobile x-ray unit
Table of Contents

Advertisement

Mobile X-ray Unit
Operation
Maximum Power kW
(Refer to
Identification Label)
Power Line Operation
Power Line Operation
(Depends on Region Con-
figuration)
Power Line Cable
Maximum Input Power
Operation independent
from mains supply
(Stand-Alone)
Battery Capacity
Radiation Output
Accuracy
(Reproducibility
related to loading factors)
Maximum Symmetrical
Radiation Field
Maximum Heat Output
Storage / Transport
Environmental
Conditions
Operating
Environmental
Conditions
138
20 kW
Automatic Line Compensation
Connection to standard outlets with GND that complies with local regulations
The Power Line Installation should be provided with a Differential of 30 mA Sensitivity
Power Line Impedance must be less than the maximum indicated value:
300 m for 100 V~, 1  for 110 V~, 2.5  for 230 V~, 2.6  for 240 V~
Optimized Battery Management for extended Battery life.
The required time for the Batteries to be 100% charged is approximately: 8 hours,
(80% of the total charge is available after 4 hours charging; approximately 20% per hour).
Total Energy storage Capacity: 5760 Wh
With the Batteries fully charged and disconnected from the mains,
the Mobile system can be in continuous movement during approximately 25 km, at 5.5 km/h
Once the exposure capacity is exhausted, the system motion is enabled for 1 km.
The Mobile Unit in Stand-Alone (disconnected from the mains) will be 100% discharged from full charge
Measured at 75 kV: 220 mm in "X" axis and 240 mm in "Y" axis.
Measured at 125 kV: 210 mm in "X" axis and 250 mm in "Y" axis.
(Test performed at a distance from the Focal Spot of 1200 mm, in accordance with IEC 60806:1984).
Relative Humidity range of 20% to 90%
Atmospheric Pressure range of 700 hPa to 1060 hPa
(the recommended temperature for a longer life cycle of batteries is around 22
Relative Humidity (no condensing) range of 30% to 75%
Atmospheric Pressure range of 700 hPa to 1060 hPa
32 kW
40 kW
100--240 V~ -- Single-Phase 50/60 Hz
10% V~
¦
Total Cable Length: 5 m
Usable Cable Length: 4.75 m
1.2 kVA
Standard
Charge Capacity: 15
Ah.
in approximately: 11 hours.
C.V. (Coefficient of variation)
260 W (1130 BTU/h)
o
Temperature range of --10
C to 40
o
Temperature range of 10
C to 35
50 kW
0.05
o
C
o
C
o
C)
OM-0481R1