Mobile X-ray Unit
Operation
ID
300002
Positioner configuration could not be loaded
300003
Switches configuration could not be loaded
300006
Generator configuration could not be loaded
300007
Workstations configuration could not be loaded
300008
Image receptors configuration could not be loaded
300009
Positioner disconnected
300010
Console disconnected
300011
Generator disconnected
300012
Workstation mismatch for Generator and Positioner
Active Procedure mismatch for Generator and
300013
Positioner
300014
Positioner configuration could not be loaded
300015
Configuration changes could not be saved
300016
Initializing communications ...
300019
Generator working in Service mode
300020
Positioner working in Service mode
300021
Could not verify Service access
300024
Usability settings configuration could not be loaded
300025
Layout settings calibration data could not be loaded
300029
Snapshots in progress
The exposure settings have been modified during
calibration. Please make sure to recover the
300031
exposure settings from the Image System
application.
126
DESCRIPTION
TYPE
Error
Error
Call Service
Call Service.
Error
Call Service.
If the error appears during start- -up, please wait. If it is
not the case, restart the System and try to repeat the
Error
operation. If the error persists, call Service.
Restart the System and try to repeat the operation. If
Restart the System and try to repeat the operation. If
Error
Error
the error persists, call Service
Try to select another workstation or another
radiographic procedure.
Error
If the error persists, restart the System and try to
repeat the operation.
If the error persists, call Service
Error
Call Service.
Error
- -
If message does not disappear after the start- -up is
Information
finished, call Service
Warning
Please restart the System
Warning
Please restart the System
Warning
- -
Error
Error
Call Service.
Call Service
Inhibit
If the message does not disappear in a period of time,
Exposure
please restart the System.
Inhibit
Recover the exposure settings from the Image System
Exposure
application.
USER HELPTEXT
OM-0481R1