Mobile X-ray Unit
Operation
ID
293034
Motion Control: Stop due to Battery charging
Motion Control: Stop due to Deadman pressed at
293035
start up
Motion Control: Stop due to Handle bar pressed at
293036
start up
Motion Control: Stop due to Header key pressed at
293037
start up
Motion Control: Stop due to Force Sensors
293038
Calibration Fault
293051
Motion Control: Header key not working properly
294001
294002
294003
Internal error located in the head.
294004
294005
294014
294015
System communication error located in the head.
S t
i
294016
294017
124
DESCRIPTION
ti
l
t d i th h
TYPE
Unplugged and plug the System again.
Information
If failure persists, restart the System.
If the error persists, call Service.
Release and push the deadman.
If failure persists, push the Handlebar.
Information
If failure persists, restart the System and try to repeat
the operation.
If the error persists, call Service.
Release the Handlebar and push it again.
If failure persists, restart the System and try to repeat
Information
the operation.
If the error persists, call Service.
Release the Header key pressed.
If fine positioning is desired, press the desired key
again.
Information
If failure persists, restart the System.
If the error persists, call Service.
Please, verify the force sensors connection.
Calibrate the force sensors again.
Information
Restart the unit and if error persists, call Service.
Warning
Restart System. If error persists call Service.
Restart the System and try to repeat the operation.
Error
If the error persists, call Service
d
USER HELPTEXT
p
,
OM-0481R1