Konica Minolta mKDR Xpress Operation page 127

Digital mobile x-ray unit
Table of Contents

Advertisement

ID
DESCRIPTION
Invalid configuration parameters according to the
290103
license installed.
290104
System will power off: Incorrect system in license
System will power off: Incorrect startup mode in
290105
license
290106
Updating license client.
290200
Collision detected
CRITICAL BATTERY LEVEL. Operation is not
290201
allowed. Please, connect the system to a power
source. Time remaining before shutdown 00:{0}:{1}
Battery Very Low, please connect the system to
290202
power supply.
Battery Low, please connect the system to power
290203
supply
290205
Collimator Apertures have not been calibrated
290206
Focal Skin Distance has not been calibrated
290207
Head- -assembly handles pressed during booting up
290208
Head- -assembly buttons pressed during booting up
290300
System communication error located in the Head.
System communication error located in the smart
290301
on/off.
290302
System communication error located in the RFID.
System communication error located in the motion
290303
control.
The system has not been used for long time and
will shutdown in 00:{0}:{1} unless you connect it to
290400
power source or click here.
290401
System is powering off. Please wait...
OM-0481R1
TYPE
Error
Call Service.
Information
- -
Inhibit
Movement
Connect the System to mains.
Warning
If the error persists, call service
Warning
- -
Information
Collimator Blades calibration is needed. Call Service.
Focal Skin Distance sensor calibration is needed. Call
Service.
Warning
Warning
Restart the System and try to repeat the operation.
If the error persists, call Service.
Error
Error
Warning
- -
Information
Mobile X-ray Unit
Operation
USER HELPTEXT