Appendix C
If the Operator presses on Accept button or performs any system action
(touches on the RAD Screen, activates any of the movement controls,
presses the handswitch/remote control (option) or connects the system
to the mains), within the configured time; it will delay the "Time Out
Inactivity Warning" for another 15 minutes.
If the Operator does not respond within the countdown time of 15
minutes, the system will automatically shut down; i.e., the system will be
completely turned OFF after 30 minutes without activity.
" REMOTE HANDSWITCH SETTINGS
Timeout Remote Handswitch Out of Holder.
Default configured time: 0.
To configure the time (in minutes) that the system will take to display an
informational message if the IR Remote Control is not placed in its holder.
Beeping Remote Handswitch.
Disabled by default.
Enable this radio button to add a discontinuous beeping to the previous
informational message each time the timeout specified in the "Timeout Remote
Handswitch Out of Holder" field is exceeded.
" FINE POSITIONING BUTTONS SETTINGS
Timeout: Seconds.
Default configured time: 5.
Allows to set the time it will take for the Unit to stop automatically when a
displacement is being performed using the Fine Positioning buttons.
Mobile X-ray Unit
Operation
C-15