Konica Minolta mKDR Xpress Operation page 42

Digital mobile x-ray unit
Table of Contents

Advertisement

Mobile X-ray Unit
Operation
GUIDANCE AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC IMMUNITY
This X-ray System is intended for use in an electromagnetic environment specified below.
The customer or user of this X-ray System should assure that it is used in such an environment.
Immunity Test
Radiated RF EM fields
IEC 61000-4-3
Proximity fields from RF
y
wireless Communications
i l
C
i
ti
equipment
IEC 61000-4-3
Conducted disturbances
induced by RF fields
induced by RF fields
IEC 61000-4-6
NOTE - - The ISM (Industrial, Scientific and Medical) bands between 0.15 MHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz
to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
The amateur radio bands between 0.15 MHz and 80 MHz are 1.8 MHz to 2.0 MHz; 3.5 MHz to 4.0 MHz; 5.3 MHz to 5.4 MHz; 7 MHz
to 7.3 MHz; 10.1 MHz to 10.15 MHz; 14 MHz to 14.2 MHz; 18.07 MHz to 18.17 MHz; 21.0 MHz to 21.4 MHz; 24.89 MHz to 24.99 MHz;
28.0 MHz to 29.7 MHz; and 50.0 MHz to 54.0 MHz.
34
(IEC 60601-1-2:2014)
IEC 60601-1-2:2014
Test Level
3 V/m
from 80 MHz to 2.7 GHz
(80% AM at 1 kHz)
Refer to next table
"IMMUNITY
"IMMUNITY
REQUIREMENTS FOR RF
WIRELESS
COMMUNICATIONS
EQUIPMENT"
3 Vrms
from 150 kHz to 80 Mhz
6 Vrms in ISM bands
from 150 kHz to 80 MHz
(80% AM at 1 kHz)
Compliance Level
3 V/m
from 80 MHz to 2.7 GHz
(80% AM at 1 kHz)
Refer to next table
"IMMUNITY
"IMMUNITY
REQUIREMENTS FOR RF
WIRELESS
COMMUNICATIONS
EQUIPMENT"
3 Vrms
from 150 kHz to 80 Mhz
6 Vrms in ISM bands
from 150 kHz to 80 MHz
(80% AM at 1 kHz)
Electromagnetic
environment - guidance
Portable
RF
communications
equipment (including peripherals such
as
antenna
cables
and
antennas) should be used no closer
than 30 cm to any part of the equipment,
including
cables
specified
manufacturer. Otherwise, degradation
of the performance of this equipment
p
could result.
could result
OM-0481R1
external
by
q p