Konica Minolta mKDR Xpress Operation page 130

Digital mobile x-ray unit
Table of Contents

Advertisement

Mobile X-ray Unit
Operation
ID
291050
291051
Internal error located in the battery monitor
Internal error located in the battery monitor.
291052
291053
Battery below critically low, shutting down the
291100
system.
291101
Download not performed due to a hardware failure
292001
292002
292003
Internal error located in the RFID.
292004
292005
292014
292015
System communication error located in the RFID
System communication error located in the RFID.
292016
292017
292023
Internal error located in the RFID.
292024
RFID System Communication Error
292040
RFID Peripheral Communication Error
RFID Peripheral Communication Error
292041
292100
RFID reset by WDT
292101
RFID not available
122
DESCRIPTION
TYPE
Restart the System and try to repeat the operation.
y
Error
Error
If the error persists, call Service.
Warning
-
Information
Download not performed due to a hardware failure.
Restart the System and try to repeat the operation.
Error
If the error persists, call Service.
If the error persists, call Service.
Restart the System and try to repeat the operation.
y
Error
Error
If the error persists, call Service.
Information
No user action.
Information
Check RFID device.
USER HELPTEXT
y
p
p
y
p
p
OM-0481R1

Advertisement

Table of Contents
loading