Konica Minolta mKDR Xpress Operation page 116

Digital mobile x-ray unit
Table of Contents

Advertisement

Mobile X-ray Unit
Operation
ID
100250
Communication lost with UARC Device
100251
Communication lost with Power Control Device
Timeout for receiving the Ready State from Power
100252
Control has elapsed
Timeout for receiving the Synch State from Power
100253
Control has elapsed
100254
Power Control Protocol Version mismatch
100255
UARC Protocol Version mismatch
100256
Starter is not supported by License
100257
High Speed is not supported by License
100258
FPGA running with Golden Firmware
100803
Activated Procedure is not allowed
100930
Tube file wrong format
Current Image Receptor Synchronization Type is
100941
not supported. Please, select other workstation.
100942
Preheating Filament
100944
Maximum Preparation Time Reached.
101008
INTERLOCK_2
101015
Generator has not been calibrated yet
101017
Configurated Tube is different to Calibrated Tube
101018
Exposure does not allowed in this Desktop
101019
The system has to be rebooted for being upgraded
101021
Inhibit Xray due to system communication lost
108
DESCRIPTION
TYPE
Press Accept.
Error
Error
If the error persists restart the System
If the error persists, restart the System.
If the error persists, call Service.
Press Accept. Release the PREP and EXP orders
from the Handswitch and try the exposure again.
from the Handswitch and try the exposure again.
Error
Error
If the error persists, call Service.
Restart the System.
y
Error
Error
If the error persists, call Service.
Error
Call Service.
Error
Error
- -
Inhibit
Change the exposure values or wait for the X- -ray
Exposure
Tube to cool.
Error
- -
Information
Wait for 10 seconds.
Information
Remove preparation and exposure order.
Inhibit
Exposure
Information
Information
- -
Inhibit
Exposure
Information
Press Accept and wait for the System to restart
Inhibit
- -
Exposure
USER HELPTEXT
OM-0481R1