2.10 Quantitative Information; Functional Tests Performed To Obtain The Quantitative Information - Konica Minolta mKDR Xpress Operation

Digital mobile x-ray unit
Table of Contents

Advertisement

2.10 QUANTITATIVE INFORMATION

.
Note
2.10.1

FUNCTIONAL TESTS PERFORMED TO OBTAIN THE QUANTITATIVE INFORMATION

.
Note
OM-0481R1
The following tables show the Quantitative Information associated
to
this
equipment
IEC 60601-1-3:2008 and IEC 60601-1-3:2008+AMD1:2013. This
information illustrates loading factors for image performance and
supplies Dose indication examples. Therefore, these tables are an
instance of the adjustment of Loading Factors, Focal Spot
Selection, SID and Collimator opening, which affect to the
radiation quality or to the radiation dose rate applied in normal use.
Equipment:
These functional tests have been performed with a specific
configuration of Digital Detector, maximum power X-ray Tube and
Collimator. The results obtained with this configuration are
representative of the worst case within the different configurations
of the unit.
Instrumentation used:
Dosimeters:
IBA KermaX 120-131 MIC CAN
G
Unfors Xi R/F
G
Thermohygrometer Testo 608-H2.
Water Phantom made of Polymethyl-methacrylate (PMMA) layers:
25 cm x 25 cm x 15 cm.
Test details:
The measurements were made using the most common APR
configurations performed with this unit.
according
with
the
Mobile X-ray Unit
Operation
Standard
37