System Overview
Optical Components
Each optical filter mount is labeled with the corresponding laser and band-pass information. See
Figure
1.5.
Figure 1.5 Optical Filter Mount Labeled with the Band-Pass Information
The top of each optical filter mount has two marks. The color of the dot
the laser. See
band-pass filter. See
Figure 1.6 Optical Filter Mount (Top)
b
c
1.
. Indicates corresponding laser color: Blue indicates a 488 nm laser; Red indicates a
638 nm laser; Violet indicates a 405 nm laser; Yellow indicates a 561 nm laser; Pink indicates
a 808 nm laser; White indicates either 375 nm or 355 nm.
2.
indicated numerically on the lateral side of the mount.
Band-pass filters are used to transmit fluorescence within a specific range of wavelength. These
ranges are designed to measure fluorescence from fluorochromes such as those listed in
(with red and violet laser upgrades installed). You can change the optical filters according to your
detector configuration. There is no need to realign the optical system when the filters are changed.
1-6
Figure
1.6. The color of the line
Figure
1.6.
. Indicates the band-pass wavelength ranges; the midpoint of the band-pass is
indicates the wavelength range of the optical
(2)
indicates the color of
(1)
Table
1.1.
B49006AS