Standardize your laser target values in the QC Experiment. Refer to
CHAPTER 5, Instrument Quality Control and
NOTE
Disabled lasers are marked Laser XXX is disabled in the QC screen and do not provide laser power
values.

Adjusting the Gain

While the instrument is in use, the signal value can be increased or decreased by adjusting the
instrument's gain configuration.
1
Select
2
Select the
Select or edit the instrument's default gain settings using one of the following methods:
• Edit the gain settings and select
• Select
• Select
NOTE
In cases where you do not specify your own default parameters, the QC settings and default
settings are identical.
B49006AS
on the left side of the screen. The Acq. Setting window appears.
tab in the Acq. Setting window.
Gain
to return to your saved default settings.
Default
to use the instrument's QC settings.
QC Gain
Data Acquisition and Sample Analysis
Standardization.
to create a new default setting.
Set as Default
Configuring Acquisition Settings
Standardization
in
6-41