Preparing the QC Sample [With Plate Loader]

Required Materials

The following materials are required to complete the QC process:
• CytoFLEX Daily QC Fluorospheres or CytoFLEX Ready to Use Daily QC Fluorospheres
• CytoFLEX Daily IR QC Fluorospheres (for systems configured with an IR laser)
• CytoFLEX Sheath Fluid or another nonionic sheath fluid
• Standard 96-well plate
— 96-well flat-bottom
— 96-well V-bottom
— 96-well U-bottom
• 96-well deep well plate
— 96-well V-bottom
— 96-well U-bottom
• Vortexer

Preparation Process CytoFLEX Daily QC Fluorospheres

IMPORTANT
CytoFLEX Daily IR QC Fluorospheres, refer to
Process
1
Take one 96-well plate and record the QC sample well position.
2
Use the vortexer or shake vigorously to thoroughly mix the bottle of CytoFLEX Daily QC
Fluorospheres.
IMPORTANT
3
Add one drop of CytoFLEX Daily QC Fluorospheres to the sample well.
4
Add 200 L of deionized water to the sample well.
5
Place the well plate in a dark location at 2-8 C until ready to load the well plate into the
instrument for QC.
NOTE
Well plates containing diluted CytoFLEX Daily QC Fluorospheres should be stored sealed in a dark
location at 2-8 C for up to 5 days.
B49006AS
For preparation instructions for the CytoFLEX Ready to Use Daily QC Fluorospheres or the
or
CytoFLEX Daily IR QC Fluorospheres Preparation
Do not overfill the sample well.