Adjusting Autogate Movement and Extent
Movement — The distance an autogate can move to find the target population.
To adjust movement, right-click an autogate and drag the Movement handle in the auto gate menu
left and right.
NOTE
The default value setting for movement is 20 units. The minimum value setting for movement is 0
units and the maximum value setting for movement is 100 units.
If a target population is consistently in the same location, movement is not needed. However, if a
target population is periodically missing from some samples, or events are rare, movement can be
used to move the gate within a certain percentage of its axis to capture the correct population.
Refer to
Figure 6.5
example of the maximum movement setting.
Figure 6.5 Movement - Default Setting
B49006AS
for an example of the default movement setting. Refer to
Data Acquisition and Sample Analysis
Creating an Experiment
Figure 6.6
for an
6
6-31