Sampling And Collecting Data [With Plate Loader] - Beckman Coulter CytoFLEX Instructions For Use Manual

Flow cytometers
Hide thumbs Also See for CytoFLEX:
Table of Contents

Advertisement

Data Acquisition and Sample Analysis
Load Sample and Record Data
Append data to existing file.
NOTE
When you select a tube that only contains acquired data, as indicated by the blue tube
test tube section of the screen, the following message appears:
Create new tube.
Overwrite existing data.
Append data to existing file.
13
Repeat steps
NOTE
If the rate suddenly appears to drop, check to see if the sample has run dry or the sample probe
is clogged. Any time the sample probe becomes clogged, immediately select Stop to unload the
sample. Then select Backflush to clean the sample probe. Refer to
Cleaning Procedures
contact

Sampling and Collecting Data [With Plate Loader]

IMPORTANT
1
To run a single well:
a. Select the well in the Plate window.
b. Select
NOTE
c. Select
d. Select
To run a set of wells:
a. Select the desired wells.
6-58
Adds new data to the existing data.
Creates a new tube in the test tube section of the screen for the data.
Overwrites the current tube data with new data.
Adds new data to the existing data.
1-12
until all sample tube data required for testing has been collected.
to flush out the sample probe. If you are still unable to clear the sample probe,
us.
Ensure the sample plate is loaded properly before acquiring the samples.
to prompt the system to begin sample aspiration.
Run
Acquisition settings can be adjusted during acquisition.
to save the data.
Record
to prompt the plate loader to eject.
Eject
in the
Daily Clean
in
CHAPTER 11,
B49006AS