Beckman Coulter CytoFLEX Instructions For Use Manual page 209

Flow cytometers
Hide thumbs Also See for CytoFLEX:
Table of Contents

Advertisement

6
Set the channels to be standardized.
NOTE
The contents of the channel, laser, and filter column come from the current detector configuration
setting. Refer to
and Sample
NOTE
Do not set the median tolerance range any lower than 5%.
NOTE
FSC is a required channel.
7
Enter the target median values saved in Step
Or copy the median values from the previously exported CSV file and paste into the
corresponding median column.
NOTE
Verify that the target values are entered correctly.
8
Select
OK
The saved results display in the Standardization Target Library window. This item is ready to
be run through the Standardization experiment. Refer to
CHAPTER 6, Data Acquisition and Sample
9
Select
B49006AS
Verifying, Selecting, Editing, and Creating Detector Configuration
Analysis.
to save the target value.
to exit the Standardization Target Library window.
Instrument Quality Control and Standardization
14
into the corresponding channels.
Performing the Standardization
Analysis.
5
Standardization
in
Data Acquisition
in
5-39