Beckman Coulter CytoFLEX Instructions For Use Manual page 183

Flow cytometers
Hide thumbs Also See for CytoFLEX:
Table of Contents

Advertisement

4
Select
Start QC/Standardization
[CytoFLEX LX Shown]
Ensure that the QC bead lot number is selectable in the Lot No. drop down menu. If the lot
number is not selectable, refer to
lot number.
5
Select
Initialize
6
Insert the prepared QC sample tube (see
Preparation
holder.
7
Select
Start
Completed processes appear on the left. Plots appear on the right. The QC experiment
sequentially detects the system configuration, laser power, laser delay, signal strength, and
coefficient of variation.
B49006AS
in the QC/Standardization menu to access the QC experiment.
Importing Lot-Specific Target
.
Process, or
CytoFLEX Daily QC Fluorospheres Preparation
to load the sample and begin to run the QC procedure.
Instrument Quality Control and Standardization
Values, then select the proper
CytoFLEX Ready to Use Daily QC Fluorospheres
Quality Control
Process) into the tube
5-13
5

Advertisement

Table of Contents
loading