Laser Setting Window [Cytoflex Lx Shown]; Target Power Ranges In The Laser Setting Screen - Beckman Coulter CytoFLEX Instructions For Use Manual

Flow cytometers
Hide thumbs Also See for CytoFLEX:
Table of Contents

Advertisement

Figure 6.9 Laser Setting Window [CytoFLEX LX shown]
1. Enable/Disable: Enables or disables the laser.
2. Target Power (mW): Used to change the laser target power.
NOTE
±1 mW tolerance. Refer to
screen.
NOTE
(refer to
3.
IMPORTANT
4.
laser power in the Actual Power (mW) column of the Laser Setting window. Refer to
of laser ranges.
Table 6.1 Target Power Ranges in the Laser Setting Screen
Laser
355
375
405
488
561
638
808
B49006AS
Laser target power can only be adjusted on the CytoFLEX LX system. The power detector has
Table 6.1
Refer to the target powers listed for each laser in the QC reports area of the QC Report Screen
Figure
2.2) for range limits.
: Sets the laser target power setting.
The Actual Power readings are +/- 1 mW.
: Reads the current laser power before the flow cell assembly and displays the current
Min Power (mW) Max Power (mW)
19
21
51
69
71
119
41
59
21
39
41
59
51
69
Data Acquisition and Sample Analysis
for the target power ranges allowed in the Laser Setting
Configuring Acquisition Settings
Table 6.1
for a list
6-39
6

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cytoflex sCytoflex lx

Table of Contents