8 Horizontal Arm Brake And Cable Assembly Installation - GE AMX-4+ Technical Manual

Hide thumbs Also See for AMX-4+:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 5
3 8
Horizontal Arm Brake and
Cable Assembly Installation
1. Mount the brake and cable assembly on the inner tube.
2. Secure the brake and cable assembly bracket to the inner tube with two
capscrews. See Illustration 3-6.
3. Carefully slide the brake and cable assembly, and the inner tube into the center
tube.
4. Replace the inner shaft, two spacers and two bearings. Secure with two retaining
rings.
5. Place center and inner tubes in an upside down position.
6. Replace the two brake capscrews securing the brake & cable assembly to the rear
of the center tube.
7. Secure the connector bracket and ground wire to the center tube with two or
three screws (depends on unit you have).
8. Replace mounting plate capscrew holding brake and cable assembly to the center
tube.
9. Replace the latch block and secure it with four screws to center tube.
10. Place the outer tube in an upside down position.
11. Slide the center and inner tubes into the outer tube.
12. Replace the two stop block capscrews and secure the brake and cable assembly
to the end of the outer tube.
13. Connect Brake Cable.
14. Extend Horizontal Arm until clearance hole on underside is exposed.
15. Replace cable clamp.
16. Return arm to upright position.
17. Replace horizontal arm assembly. Refer to Section 3-6.