5 Counterweight; Remove Counterweight - GE AMX-4+ Technical Manual

Hide thumbs Also See for AMX-4+:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 5
4 5
Counterweight
4 5 1

Remove Counterweight

WARNING
ILLUSTRATION 4-9
COUNTERWEIGHT
1. Remove vertical column. Refer to Section 4-1.
2. Remove the main, safety and follower cables. Refer to Section 4-3.
3. Remove upper pulley and brake assembly. Refer to Section 4 4.
PERSONAL INJURY!
THE COUNTERWEIGHT WEIGHS APPROXIMATELY 100 POUNDS (45
KG). USE THE PROPER SAFETY PRECAUTIONS WHEN HANDLING
THE COUNTERWEIGHTS.
4. Slide counterweight from vertical column.
5. Remove top cable retainer clip from counterweight assembly by removing two
slotted hex washer head screws. See Illustration 4-9.
STUD
ENDS
SCREW
RETAINER
CLIP
RETAINER
CLIP
SCREW
4-16
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
MAIN AND SAFETY CABLES
BALL
ENDS
AMX-4+ SERVICE
COUNTERWEIGHT
FOLLOWER
CABLE

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

2169360 series2236420 series2275938 series