Main Cable - GE AMX-4+ Technical Manual

Hide thumbs Also See for AMX-4+:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 5
4 3 1

Main Cable

CAUTION
ILLUSTRATION 4-4
COUNTERWEIGHT AND CABLES
UPPER
PULLEY
MAIN
CABLE
SAFETY
CABLE
CABLE
STUD
END
FRONT
OF UNIT
1. Loosen locking nut on top of stud end of main cable. Main cable is in front of
follower cable.
After next step the counterweight is held only by the safety cable.
2. Remove holding nut on bottom of stud end of main cable.
3. Remove cable stud from carriage cable mounting strut.
4. Move horizontal arm to bottom of column.
D
This raises the counterweight to the top of the column.
5. Remove counterweight trim weights (if present). See Illustration 4-4.
CABLE
BALL
END
RETAINER
CLIP
4-5
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
SCREW
AMX-4+ SERVICE
TRIM
WEIGHT
COUNTERWEIGHT