GE MEDICAL SYSTEMS
REV 5
4 6
Carriage Assembly
4 6 1
Remove Carriage
ILLUSTRATION 4-10
CARRIAGE ASSEMBLY
4 6 2
Install Carriage
4 6 3
Remove Carriage Bearings
1. Remove upper pulley and brake assembly. Refer to Section 4-5.
2. Remove follower cable end from carriage strut by loosening its holding nut. See
Illustration 4-10.
3. Slide carriage to top of column and remove it.
MAIN
CABLE
CABLE
STUD
END
HOLDING
NUT
FOLLOWER
CABLE
1. Slide carriage assembly into top of column. Check that horizontal arm mounting
plate is at top of carriage.
2. Install pulley and brake assembly. Refer to Section 4-6.
3. Place follower cable end into carriage strut and secure in place by tightening its
holding nut. See Illustration 4-10.
1. Remove carriage assembly. Refer to Section 4 6 1.
2. Remove four hex head shoulder bolts securing four forward facing bearings to
side flanges on carriage. See Illustration 4-11.
3. Remove carriage side facing bearing shoulder bolts and bearings from carriage.
4-19
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
SAFETY
CABLE
AMX-4+ SERVICE