GE MEDICAL SYSTEMS
REV 5
ILLUSTRATION 6-3
BATTERY CONNECTIONS
BATTERY SEPARATOR PAD
TYPICAL (4 PLACES)
SEE NOTE 3
CABLE 1
SENSE LEAD
CABLE 1
CABLE 6
CABLE 8
CABLE 7
SENSE LEAD
5
SEE
NOTE 2
CABLE 1
CABLE 1
SENSE LEAD
NOTE 1: THERE ARE NO INSULATORS #3 & #7.
NOTE 2: DON'T CONFUSE INSULATOR #4 WITH INSULATOR #1. INSULATOR #4 HAS TWO
NON-FUNCTIONAL HOLES. INSULATOR #1 HAVE TWO CUP-SHAPED BUMPS WITH NO HOLES.
INSULATOR #1 MUST BE ORIENTED WITH BUMPS POINTING AWAY FROM YOU.
NOTE 3: ADD BATTERY SEPARATOR PADS BETWEEN BATTERIES 1 & 3, 2 & 4, 5 & 6, AND 7 & 8.
DON'T REMOVE PAPER BACKING FROM BATTERY SEPARATOR PADS. ORIENT PADS WITH PAPER
BACKING FACING UP .
6-4
(MODEL 2169360, 2236420 & 2275938 SERIES)
CABLE 7
SENSE LEAD
CABLE 7 (2)
SEE
NOTE 2
CABLE 3
CABLE 3
SENSE LEAD
SEE
NOTE 2
CABLE 2
SENSE LEAD
AMX-4+ SERVICE
DIRECTION 2173225-100
CABLE 4
SENSE LEAD
CABLE 4 TO
CIRCUIT BREAKER
CABLE 6
SENSE LEAD
BATTERY TEST
TERMINAL STRIP
CABLE 5
CABLE 5
SENSE LEAD
SEE
NOTE 2
CABLE 2
TO GROUND STUD